Insight 102

Two EU Dates Outside the MDR: Battery Removability (18 February 2027) and the Machinery Regulation (20 January 2027) for US Device and Drug-Delivery Makers

This article provides general information and is not legal advice.

Abstract: From 18 February 2027 end-user replacement is the default for portable batteries in newly placed products. Qualifying on-body delivery systems and wearables may instead allow independent-professional replacement. From 20 January 2027 the Machinery Regulation supplies the safety requirements referenced by the MDR and IVDR where the relevant hazard and specificity tests are met.
Plain Language Summary

In early 2027 two sets of EU product rules affect medical devices and drug-delivery products. One generally requires users to be able to remove and replace portable batteries and permits professional replacement or a narrower exception in certain cases. The other changes which machinery safety requirements the medical-device rules refer to. Which rules apply depends on the product and the hazards it presents.

Contents
  1. Why Two Product Laws Reach a CE-Marked Device
  2. 18 February 2027: The End User Must Be Able to Remove the Battery
  3. Professional Replacement and the Permanent-Connection Derogation
  4. 20 January 2027: Machinery Requirements Arrive Through the MDR
  5. What the Dates Do to Import and Distribution Terms
  6. Strategic Considerations for US Device and Drug-Delivery Makers

A US device team plans a European launch against two files, the one that supported clearance by the Food and Drug Administration and the one built for the Medical Device Regulation. Two dates in early 2027 come from neither. On 20 January 2027 the Machinery Regulation replaces the directive whose safety requirements the device rules reference for products that also qualify as machinery, and on 18 February 2027 the Batteries Regulation requires that the end user be able to remove and replace the portable battery of a product placed on the Union market, subject to defined derogations. A design that requires a different form of replacement needs to meet the applicable conditions.

1. Why Two Product Laws Reach a CE-Marked Device

The Batteries Regulation applies to every category of battery, including batteries incorporated into products, whatever the product is. Art. 1(5) of the Batteries Regulation excludes batteries incorporated in or designed for certain military or space equipment. It contains no general exclusion for medical devices.1Regulation (EU) 2023/1542 (Batteries Regulation), Art. 1(3), Art. 1(5) and Art. 96(2).

Machinery law reaches devices through the device rules themselves. Under Art. 1(12) MDR, a device that is also machinery must, where a relevant hazard exists, meet the essential health and safety requirements of the Machinery Directive to the extent that they are more specific than the general safety and performance requirements of the device rules, and Art. 1(6) IVDR applies the same test to an in vitro diagnostic device that is also machinery. Machinery is defined broadly as an assembly with a drive system other than directly applied human effort and with linked parts of which at least one moves. A surgical robot or a powered patient lift illustrates why the definition can matter. A pump or analyzer still calls for a check of its particular construction and the hazards involved.2Regulation (EU) 2017/745 (MDR), Art. 1(12); Regulation (EU) 2017/746 (IVDR), Art. 1(6).

Both references still name Directive 2006/42/EC. Art. 51(2) of the Machinery Regulation repeals that Directive with effect from 20 January 2027 and provides that references to it are to be construed as references to the Regulation and read with a correlation table. From that date read the device-law cross-references to the Directive as corresponding references to the Regulation, including the relevant requirements in its Annex III.3Regulation (EU) 2023/1230 (Machinery Regulation), Art. 3, Art. 51(2) and Art. 54.

2. 18 February 2027: The End User Must Be Able to Remove the Battery

Art. 11(1) of the Batteries Regulation obliges any person that places on the market a product incorporating a portable battery to ensure that the battery is readily removable and replaceable by the end user at any time during the life of the product. The duty concerns whole batteries. A portable battery is sealed, weighs five kilograms or less, is not designed specifically for industrial use and is neither an electric vehicle, light means of transport nor starting, lighting and ignition battery. The battery is readily removable if it can be taken out with commercially available tools, without specialized tools unless they are supplied free with the product, and without proprietary tools, heat or solvents. The product must come with instructions and safety information on removal and replacement, which must also be permanently available online.4Batteries Regulation (n 1), Art. 3(1), point (9), and Art. 11(1) and (6).

Three further duties travel with the main rule. The battery must be replaceable by a compatible battery without affecting the functioning, performance or safety of the product. Under Art. 11(7) of the Batteries Regulation it must be available as a spare part for at least five years after the last unit of the model is placed on the market, at a reasonable and non-discriminatory price. And under Art. 11(8) of the Batteries Regulation software must not be used to impede replacement with a compatible battery. The Commission's guidelines give parts pairing as an example of software that impedes replacement and recognize that software may check a replacement battery for correct functioning and safety. For a device whose manufacturer authenticates components for reasons of safety and cybersecurity, the line between the two is not drawn in the text.5Batteries Regulation (n 1), Art. 11(7) and Art. 11(8); Commission Notice C/2025/214.

The duty attaches to the act of placing a product on the market, and Union product law applies that concept to each unit and not to a product model. A design that has been on sale for years is caught for every unit first supplied in the Union from 18 February 2027. For an MDR device from a third country, the EU-established person that places it on the Union market is the importer under Art. 2(33) MDR. An EU affiliate or distributor can fill that role. For a product outside the MDR, assess who first places the product under the Batteries Regulation on the facts of that supply.

For an MDR device from a third country, the EU importer places the unit on the market. The battery duty may therefore fall on a different entity from the manufacturer that controls the design.

3. Professional Replacement and the Permanent-Connection Derogation

Art. 11(2)(b) of the Batteries Regulation allows professional medical imaging and radiotherapy devices and in vitro diagnostic medical devices to be designed for battery removal and replacement by independent professionals. The delegated act retains that provision. Devices can also qualify under another product category where its conditions are met.

The delegated act adds other qualifying categories. Commission Delegated Regulation (EU) 2026/1770, adopted on 14 July 2026 and published on 22 October 2026, replaces Art. 11(2) of the Batteries Regulation and adds further product groups to the list. Two concern this readership: on-body delivery systems specifically designed for the subcutaneous administration of medicinal products, and wearable devices whose safety, durability or water resistance may be compromised by user access to the battery and which are too small for the user to handle a replacement safely or depend on a compact sealed enclosure to maintain functional integrity. For both additions, the derogation must be necessary for the safety of the user and the appliance. The battery still has to be removable and replaceable by an independent professional. Neither the Regulation nor the delegated act says how a patch pump or injector that is discarded after a single use meets that duty.6Commission Delegated Regulation (EU) 2026/1770 on portable-battery derogations [2026] OJ L 2026/1770, Art. 1 (points (b), (c) and (g) and final sentence) and Art. 2.

Art. 11(3) of the Batteries Regulation disapplies the obligations in its first paragraph where continuity of power supply is necessary and a permanent connection between product and battery is required to ensure the safety of the user and the appliance or, for products that collect and supply data as their main function, for reasons of data integrity. The Commission's guidelines treat implantable active devices, certain high-risk diagnostic devices and hearing aids as relevant to this exemption and suggest using the risk classes of the device rules as a guide. The nonbinding guidelines use these as examples and call for product-specific assessment. Art. 11(3) of the Batteries Regulation addresses only the obligations in paragraph 1. The separate spare-part and software duties in Art. 11(7) and (8) of the Batteries Regulation require their own analysis.7Batteries Regulation (n 1), Art. 11(3), (7) and (8); Commission Notice C/2025/214 (n 5), sections 1 and 4.

These categories matter for glucose sensors, connected pens and portable infusion pumps. Applying the derogations requires facts about the particular product and its hazards. The importer needs the manufacturer's assessment. Existing conformity assessment under the device rules does not by itself establish compliance with the battery requirements.

4. 20 January 2027: Machinery Requirements Arrive Through the MDR

The Machinery Regulation carries over most of the Directive's requirements and adds several that concern software. Section 1.1.9 of Annex III to the Machinery Regulation requires protection against corruption: software and data critical for safety must be identified and protected against accidental or intentional corruption, and the product must collect evidence of legitimate and illegitimate interventions in that software. Section 1.2.1 of Annex III to the Machinery Regulation addresses reasonably foreseeable malicious attempts that could cause a hazard, where appropriate to the risks. It limits changes to settings or rules that could create hazards, including during learning. It also requires a log of intervention data and safety software versions uploaded after placing on the market or putting into service, enabled for five years after the upload to demonstrate conformity on a reasoned request.8Machinery Regulation (n 3), Annex III, sections 1.1.9 and 1.2.1.

Whether these requirements bind a given device depends on the filter in Art. 1(12) MDR. They apply only where a hazard relevant under machinery law exists and only to the extent that they are more specific than the device rules, which have their own requirements on software, cybersecurity and risk management. A manufacturer relying on an earlier analysis against the Directive needs to check the new requirements against its device documentation and the hazards presented by the intended use. Art. 52(1) of the Machinery Regulation protects products placed on the market in conformity with the Directive before 20 January 2027; it does not protect later units of the same model.

The Regulation has been amended since 2023, and one amendment concerns devices. The Digital Omnibus on AI of July 2026 moved machinery to the sectoral list of the AI Act and instructed the Commission to add health and safety requirements for high-risk AI systems to Annex III to the Machinery Regulation by delegated act, which are to apply by 2 August 2028. The 20 January 2027 machinery date still stands. For an AI-enabled device that is also machinery, check the device route under the AI Act and, when adopted, the future machinery safety requirements. Each depends on the product classification and the applicable hazard or high-risk criteria.9Regulation (EU) 2026/1744, Art. 3(1) and recital 42.

5. What the Dates Do to Import and Distribution Terms

An importer placing the product on the EU market bears the placing-on-the-market duty and may depend on the manufacturer's design, evidence and spare parts to fulfill it. The contract can require conformity evidence, access to a claimed derogation's safety assessment and an indemnity. Review the scope of any existing warranty of compliance with applicable law and which party can supply the material the importer needs.

The spare-part duty outlasts the commercial relationship. Five years of battery availability run from the last unit of the model placed on the market, which may be years after a distribution agreement has ended or a product line has been sold. Art. 11(7) of the Batteries Regulation addresses the person placing the product on the market. Ending the distribution agreement does not itself allocate continuing performance. The agreement can specify who supplies spare parts and keeps online instructions available when the distributor or manufacturer changes.

For machinery the contractual point lies upstream. Motors, actuators and control software are often bought from suppliers that declared conformity with the Directive, and the manufacturer's own analysis under the device rules rests on those declarations. A supply agreement that names Directive 2006/42/EC should be checked for how it deals with a successor and who provides updated conformity evidence.

6. Strategic Considerations for US Device and Drug-Delivery Makers

Both dates were fixed in 2023, and neither appears in the transition tables of the device rules. The delegated act that narrows the battery rule for on-body delivery systems and small wearables entered into force only in November 2026, three months before the rule applies, and it helps only products that fit its wording and can show the safety case it requires. Where professional replacement is unavailable, the manufacturer must establish whether the permanent-connection conditions apply or meet the end-user duty. A redesign may also require a review of its effects on device conformity.

Several questions remain open. The guidelines connect the permanent-connection derogation to risk class without saying which classes qualify, so two manufacturers of comparable devices may reach opposite conclusions and both act in good faith. A single-use delivery system with a portable battery still needs an assessment of the applicable text. If it relies on professional replacement, the manufacturer needs evidence that the battery can in fact be replaced by an independent professional. An authentication feature likewise needs a basis for ensuring safety without impeding replacement by a compatible battery. For machinery, the test of what is more specific than the device rules has no authoritative reading for the new software requirements, and a notified body reviewing a device file may read it differently from a market surveillance authority. The separate product-law simplification file, COM(2025) 504 (2025/0134(COD)), had an agreed negotiating text in June 2026. The proposal itself does not change the January and February 2027 dates.

REFERENCES

01
Regulation (EU) 2023/1542 of the European Parliament and of the Council of 12 July 2023 concerning batteries and waste batteries [2023] OJ L191/1 (Batteries Regulation), Art. 1(3), Art. 1(5), Art. 3(1), points (9) and (16), and Art. 96(2). ↩
02
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices [2017] OJ L117/1 (MDR), Art. 1(12) and Art. 2(28) and (33); Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices [2017] OJ L117/176 (IVDR), Art. 1(6). ↩
03
Regulation (EU) 2023/1230 of the European Parliament and of the Council of 14 June 2023 on machinery [2023] OJ L165/1, as corrected by Corrigendum [2023] OJ L169/35 (Machinery Regulation), Art. 3, point (1)(a), Art. 9, Art. 51(2), Art. 52(1), Art. 54 and Annex XII, repealing Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery [2006] OJ L157/24. ↩
04
Batteries Regulation (n 1), Art. 3(1), point (9), Art. 11(1), first to third subparagraphs, and Art. 11(6); Commission Notice, The ‘Blue Guide’ on the implementation of EU product rules 2022 [2022] OJ C247/1, section 2.3. ↩
05
Batteries Regulation (n 1), Art. 11(7) and Art. 11(8); Commission Notice, Commission guidelines to facilitate the harmonised application of provisions on the removability and replaceability of portable batteries and LMT batteries in Regulation (EU) 2023/1542 [2025] OJ C 2025/214, sections 4 and 5. ↩
06
Commission Delegated Regulation (EU) 2026/1770 of 14 July 2026 on supplementing Regulation (EU) 2023/1542 of the European Parliament and of the Council as regards derogations for the removability and replaceability of portable batteries [2026] OJ L 2026/1770 (22 October 2026), Art. 1 (points (b), (c) and (g) and final sentence), Art. 2 and recitals 2, 3 and 7. ↩
07
Batteries Regulation (n 1), Art. 11(3), (7) and (8) and recital 38; Commission Notice C/2025/214 (n 5), sections 1 and 4. ↩
08
Machinery Regulation (n 3), Annex III, sections 1.1.9 and 1.2.1, points (a), (d) and (f). ↩
09
Regulation (EU) 2026/1744 of the European Parliament and of the Council of 8 July 2026 amending Regulations (EU) 2024/1689, (EU) 2018/1139 and (EU) 2023/1230 as regards the simplification of the implementation of harmonised rules on artificial intelligence (Digital Omnibus on AI) [2026] OJ L 2026/1744, Art. 3(1) and recital 42. ↩

GAWEL Legal advises US and Swiss device and drug-delivery manufacturers on EU and Swiss product regulation beyond the MDR and on the supply, import and distribution agreements that allocate it.

Make an enquiry