Insight 103

FDA Clearance and Swiss Market Access for Medical Devices: The Current Rules and the Proposed Reliance Route

This article provides general information and is not legal advice.

Abstract: FDA clearance alone gives no commercial route to the Swiss market. The Federal Council has set out a simplified conformity check by independent private bodies and aims to amend the ordinances by the third quarter of 2027. The target concerns the amendments. Eligibility and the date manufacturers could first use the route remain open.
Plain Language Summary

Medical devices that are allowed on the US market cannot be sold in Switzerland on that basis alone. Switzerland plans to let private bodies use the FDA's work when they check whether a device meets Swiss requirements. The government aims to amend the rules by the third quarter of 2027. For a US manufacturer, the open questions are which products will qualify and when it can start using the new route.

Contents
  1. Current Swiss Conformity Rules
  2. The Motion, the Federal Council Decision and the Trade Declaration
  3. The Proposed Swiss Review and Its Open Conditions
  4. EU Market Access and the Swiss Timetable
  5. Strategic Considerations for US Device Makers

FDA clearance or approval can allow a device onto the US market. Swiss commercial entry still depends on the applicable Swiss conformity rules. Parliament instructed the Federal Council in 2022 to make room for devices from outside Europe. The Federal Council chose a simplified conformity check in 2025 and said in June 2026 that it intends to amend the relevant ordinances by the third quarter of 2027. A US manufacturer deciding whether to wait still has no confirmed date on which it could place its device on the Swiss market under that approach.

1. Current Swiss Conformity Rules

For ordinary commercial placement, Art. 13(1) MepV requires the Swiss conformity marking or the EU Medical Device Regulation (MDR) marking. Art. 13(2) MepV lists cases in which no mark may be affixed, including custom-made devices. Under Art. 23 MepV, the assessment follows the MDR. Where a certificate is required, Art. 25(4) MepV treats certificates from EU-designated bodies established in the EU or the European Economic Area as equivalent to Swiss certificates on stated conditions. In vitro diagnostics have corresponding rules in Art. 12 IvDV, Art. 19 IvDV and Art. 21(4) IvDV.1MepV, Art. 13, Art. 23 and Art. 25(4). IvDV, Art. 12, Art. 19 and Art. 21(4).

For a US manufacturer, the current route also requires a written mandate for an authorized representative under Art. 51(1) MepV or Art. 44(1) IvDV. Swissmedic accepts a representative established in Switzerland or Liechtenstein. The importer checks the marking where applicable, the declaration of conformity and the mandate under Art. 53(1) MepV or Art. 46(1) IvDV. The representative and importer register with Swissmedic before first placing a device on the market under Art. 55(1) MepV or Art. 48(1) IvDV. Product registration is a separate obligation under Art. 17(5) MepV and Art. 16(5) IvDV. For products subject to registration and placed on the Swiss market from 1 July 2026, the general transitional deadline is 31 December 2026. Registration is immediate if a serious incident, field safety corrective action or trend must be reported. The analysis of Swiss authorized representative mandates describes what the representative takes on.2MepV (n 1), Art. 17(5), Art. 51(1), Art. 53(1), Art. 55(1), Art. 108(2) and Art. 108(3). IvDV (n 1), Art. 16(5), Art. 44(1), Art. 46(1), Art. 48(1), Art. 90(2) and Art. 90(2bis). Swissmedic, Obligations Economic Operators CH.

Two limited exceptions matter for a device that has not undergone the applicable conformity assessment. Under Art. 22(1)(a) MepV, Swissmedic may authorize a specific device on reasoned application where its use is in the interest of public health or patient safety or health. Art. 22(2) MepV permits placement and use of individual devices without that authorization if five conditions are met. They include a life-threatening condition or permanent impairment of a bodily function, the absence of a conforming device for the intended purpose, use only by a medical professional on an individual patient, and informed consent. Art. 18 IvDV has corresponding exceptions, with a laboratory-use condition and an exclusion for self-tests in Art. 18(2) IvDV. These are exceptional routes for specific devices. A US marketing decision alone fulfills neither provision.3MepV (n 1), Art. 22(1)(a) and Art. 22(2). IvDV (n 1), Art. 18(1)(a) and Art. 18(2).

A US manufacturer with FDA clearance alone still needs to establish Swiss conformity to launch. The proposed simplified procedure would make the FDA file useful in a separate Swiss review.

2. The Motion, the Federal Council Decision and the Trade Declaration

Parliament referred Motion 20.3211 to the Federal Council on 28 November 2022. It instructs the government to amend the legislation so that Switzerland can admit devices from non-European regulatory systems and widen its sources of supply. The motion sets no implementation date.4Motion 20.3211 (Damian Müller), text and National Council vote of 28 November 2022.

On 30 April 2025 the Federal Council set its approach. A working group from the BAG, Swissmedic, the SECO and the foreign ministry had identified differences between the Swiss and US systems. The press release named compliance with Swiss data protection law, a quality management system, a post-market surveillance plan and the possibility of requiring clinical evidence. The Council decided that independent private bodies should check compliance with those conditions through a simplified procedure that draws on the FDA's assessment. It instructed three departments to develop that approach.5Bundesrat, Medienmitteilung of 30 April 2025.

On 29 June 2026 Switzerland issued a declaration under the trade framework of 14 November 2025 between the United States, Switzerland and Liechtenstein. It states Switzerland's intention to facilitate Swiss market access for FDA-cleared or approved devices and amend the relevant ordinances by the third quarter of 2027. In a separate paragraph, Switzerland says it intends to treat US conformity assessment bodies no less favorably than its own. The declaration makes these actions conditional on the United States committing to the agreed tariff treatment for goods from Switzerland and Liechtenstein. The declaration does not create rights or obligations under international law.6Swiss Federal Council, Declaration by Switzerland of 29 June 2026.

On 17 November 2026, the federal research database listed the BAG's impact analysis for this motion as in progress, from 1 January to 31 December 2026. The BAG's medical device legislation page still said the working group was preparing the changes and the current rules remained in force. The page gave no text of proposed amendments for this route.7ARAMIS, project 0142007571, Basic data and Texts. BAG, Medical devices legislation, Implementation of motion 20.3211.

3. The Proposed Swiss Review and Its Open Conditions

The 2025 decision describes a private body checking compliance with Swiss device requirements. For medicines, Art. 13 HMG provides a different mechanism in which Swissmedic takes account of test results from a country with comparable medicines regulation, as the analysis of Swissmedic authorization after FDA approval explains. Art. 45(3) HMG and Art. 46(2)–(3) HMG give the Federal Council the power to set device requirements and conformity assessment procedures. Those provisions explain the government's plan to amend ordinances. The published decisions leave the required documents and the form of evidence of Swiss conformity open.8HMG, Art. 13, Art. 45(3) and Art. 46(2)–(3), version in force on 17 November 2026.

Under the FDA's terminology, a 510(k) clearance rests on substantial equivalence to a legally marketed predicate. A premarket approval (PMA) follows a separate safety and effectiveness review. The FDA grants a De Novo request to classify an eligible novel device without a legally marketed predicate. A device exempt from 510(k) may be marketed without a clearance if the exemption applies and the other applicable requirements are met. The Swiss declaration uses the terms cleared or approved. It does not say how Switzerland will treat De Novo orders or devices lawfully marketed without an individual FDA decision.9FDA, Medical Device Safety and the 510(k) Clearance Process, and Overview of Device Regulation.

The product file raises two further questions. In a December 2025 reply to Parliament, the Federal Council said a reviewer may need clinical evidence of a device’s performance where the FDA did not require it. It also said the activities of the private bodies would need criteria of their own. A manufacturer relying on a 510(k) will need to know when its FDA file supplies enough clinical evidence and when the reviewer will ask for more. The Council also identified compliance with Swiss data protection rules. For a connected device, the unresolved question is what evidence of compliance the body will ask the manufacturer to provide.10Federal Council response to Parliamentary Question 25.8087, 8 December 2025. Bundesrat (n 5).

The announcements leave the relationship between the two ordinances open, including whether the new route will apply to in vitro diagnostics at the same time. They leave the marking, the eligible review bodies and the administrative requirements open too. The conformity assessment body paragraph of the trade declaration supplies no list of authorized reviewers for this route. The current rules on the representative, importer, registration and vigilance remain the basis for planning until an amendment specifies how the new route will work.

4. EU Market Access and the Swiss Timetable

EU placement remains governed by the applicable EU regime. The European Commission has treated Switzerland as a third country for MDR devices since May 2021 and for in vitro diagnostics since May 2022, while Swiss law provides for recognition of EU certificates subject to its own conditions. For a US manufacturer pursuing both markets, a Swiss assessment under the proposed route would leave the EU conformity process to be completed, including a notified-body certificate where the applicable EU procedure requires one. The analysis of Bilaterals III and the Swiss mutual recognition agreement explains the treaty question.11BAG, Medical devices legislation (n 7), Pending update of the MRA. European Commission, Notice to Stakeholders, 26 May 2021.

The adopted motion instructs the Federal Council to act under Art. 120 ParlG. The trade declaration gives a target for amending the ordinances. An entry-into-force date would have to be set separately. The Vernehmlassungsgesetz requires consultation for ordinances of major importance and certain other ordinances, including those carried out to a substantial degree outside the federal administration. The ordinary minimum period is three months. The law specifies when a consultation may be waived, when the period must be extended and when urgency allows it to be shortened. The length of a consultation therefore gives manufacturers no fixed date by which a draft must appear.12ParlG and Vernehmlassungsgesetz.

5. Strategic Considerations for US Device Makers

A manufacturer that can satisfy the current Swiss conformity rules can plan commercial entry on that basis. For a manufacturer whose only path currently runs through the FDA, the immediate decision is how much Swiss preparatory work to pay for before it knows whether the device qualifies. A representative or distributor agreement signed now may still be in force when the ordinances change. Whether a new assessment would fit the existing mandate and how its cost would be allocated depend on the agreement and the eventual rules.

A manufacturer already selling in Switzerland under a certificate needs to know whether it could switch to the new route before the certificate expires. For software, the operational question can recur with every release. The FDA may review and authorize a predetermined change control plan as part of a marketing submission for an AI-enabled device. Changes specified in the authorized plan and implemented according to it can proceed without a new US marketing submission. The Swiss documents leave open how the reviewers would assess those changes. The time and cost of a Swiss check remain unsettled, and the ARAMIS end date records a study milestone rather than a date for release of findings.13FDA, August 2025 final guidance on predetermined change control plans for AI-enabled device software functions, §§ I, IV and V.D. ARAMIS (n 7).

REFERENCES

01
Medizinprodukteverordnung (MepV) vom 1. Juli 2020 (SR 812.213), Art. 13, Art. 23 and Art. 25(4). Verordnung über In-vitro-Diagnostika (IvDV) vom 4. Mai 2022 (SR 812.219), Art. 12, Art. 19 and Art. 21(4). ↩
02
MepV (n 1), Art. 17(5), Art. 51(1), Art. 53(1), Art. 55(1), Art. 108(2) and Art. 108(3). IvDV (n 1), Art. 16(5), Art. 44(1), Art. 46(1), Art. 48(1), Art. 90(2) and Art. 90(2bis). Swissmedic, Obligations Economic Operators CH (MU600_00_016, version 8.0, 26 October 2026), § 1.2. ↩
03
MepV (n 1), Art. 22(1)(a) and Art. 22(2). IvDV (n 1), Art. 18(1)(a) and Art. 18(2). ↩
04
Motion 20.3211 (Damian Müller), «Für mehr Handlungsspielraum bei der Beschaffung von Medizinprodukten zur Versorgung der Schweizer Bevölkerung», AB 2022 N, Beilagen, 114 and National Council voting record of 28 November 2022. ↩
05
Bundesrat, «Bundesrat legt Leitlinien für die Ausweitung der Versorgung mit Medizinprodukten fest», Medienmitteilung (Bern, 30 April 2025). ↩
06
Swiss Federal Council, Declaration by Switzerland on Further Action by Switzerland under the ‘Joint Statement on a Framework for a United States – Switzerland – Liechtenstein Agreement on Fair, Balanced, and Reciprocal Trade’ issued on November 14, 2025 (Berne, 29 June 2026), the conformity assessment bodies and medical devices paragraphs and the final paragraph. ↩
07
ARAMIS, project 0142007571, Basic data and Texts, record last modified 28 October 2026. Bundesamt für Gesundheit, ‘Medical devices legislation’, ‘Implementation of motion 20.3211’. ↩
08
Bundesgesetz über Arzneimittel und Medizinprodukte (Heilmittelgesetz, HMG) vom 15. Dezember 2000 (SR 812.21), Art. 13, Art. 45(3) and Art. 46(2)–(3) (version in force on 17 November 2026). ↩
09
FDA, ‘Medical Device Safety and the 510(k) Clearance Process’ and ‘Overview of Device Regulation’. ↩
10
Federal Council, response of 8 December 2025 to Parliamentary Question 25.8087 (Valérie Piller Carrard), «Quels critères pour la procédure simplifiée d’autorisation des dispositifs médicaux approuvés par la FDA ?», French response. Bundesrat (n 5). ↩
11
Bundesamt für Gesundheit (n 7), ‘Pending update of the MRA and mitigation measures’. European Commission, ‘Notice to Stakeholders: Status of the EU-Switzerland Mutual Recognition Agreement (MRA) for Medical Devices’ (26 May 2021), pp. 1–2. ↩
12
Bundesgesetz über die Bundesversammlung (Parlamentsgesetz, ParlG) vom 13. Dezember 2002 (SR 171.10), Art. 120. Bundesgesetz über das Vernehmlassungsverfahren (Vernehmlassungsgesetz, VlG) vom 18. März 2005 (SR 172.061), Art. 3(1)(d)–(e), Art. 3a and Art. 7(3)–(4). ↩
13
FDA, Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions (final guidance, Aug. 18, 2025), §§ I, IV and V.D. ARAMIS (n 7), Basic data. ↩

GAWEL Legal advises US device manufacturers on Swiss market access and structures authorized representative mandates, import arrangements and distribution agreements.